Gas sampling and conditioning technology
10-1.1-ME
The inner seals can be removed with a pointed tool (e.g. marking tool).
When refitting, the outer seals must be slipped over the cartridge into the respective seal groove.
The inner seals must be fitted as follows:
Place the seal in the adapter opening.
Push the O-ring with a blunt object on to the respective seal groove.
N O T E !
Refit the O-ring seals with care:
Do not damage O-rings and observe correct position of the seals !
The two outer O-ring seals of the adsorber cartridge must be replaced as follows:
Remove the old seals
Carefully slip the new seals over the cartridge into the appropriate grooves.
N O T E !
Refit the O-ring seals with care:
Do not damage O-rings and observe correct position of the seals !
15.7
LEAK CHECK
Tightly close the combustion air outlet.
Start the combustion air conditioning unit as specified in Chapter 13.
Monitor the inlet and outlet pressure on the pressure gauges on the front panel of the combustion
air conditioning unit.
Leaks are indicated when the specified pressures are not reached.