Gas sampling and gas conditioning technology

4-3.1-ME

Indication

Problem

Possible Cause

Check/Solution

Middle LED on
cooler is green;












LED of the LA
electronics is
green (see 4).

Cooler and
gas pump in
operation;
condensate in
sample gas
lead;

Pre-filter  clogged
up;


Pump hose faulty;


Peristaltic pump
SR25.1 faulty;


Unsufficient drying of
sample gas;

Optional liquid alarm
sensor:
Sensor has not turned
pump off.

Remove filter from condensate lead;
Pump delivering?
Change filter;
Pump not delivering?
Change pump hose (see manual peristaltic pump
SR25.1, 3-7.1-ME);
OK?
Check peristaltic pump (see manual peristaltic
pump SR25.1, 3-7.1-ME);

OK?
Check cooler (see manual ECP1000/3000, 3.1.1-
ME);




Check the LA electronics and if necessary replace.