Gas sampling and gas conditioning technology

3-3.0EX-ME

The gas to be measured is taken from the ECM-ExII gas cooler

by a gas sample probe

via a

heated sample line

and cooled down to a dew point of +5°C. The super-fine filter

located

afterwards removes solid particles. For increased operating safety of the entire system we
recommend installing a super-fine filter

with a liquid alarm sensor. If required an aerosol filter

can be installed in front of the flow meter

. The gas thus treated can now be passed into the

analyser

.



10

TECHNICAL DATA

10.1

FOR BASIC COOLER ECM-EXII WITHOUT HEAT EXCHANGER

Gas cooler series EC

version ECM-1 ExII

version ECM-2 ExII

Part No.

02K7600 (a)**

02K7610 (a)**

Number of heat exchanger
possibility

prepared for 1 stream

prepared for 2 streams

Sample outlet dewpoint

range of adjustment: +2 °C ..... +7 °C, factory setting: +5 °C

Dew point stability

at const. conditions: < ±0,1°C

Sample inlet temperature

***max. +180°C

Sample inlet dew point

***max. +80°C

Gas flow rate

***max. 250l/h

Ambient temperature

+10°C to +50°C

Storage temperature

-20 to +60°C

Total cooling capacity

40W at 10°C - 50°C ambient temperature

Ready for working

< 15 min

Power consumption

max. 200VA (start up current at 230V=2,5A; at 120V=4,5A)

Main power connection

230V/115V, 50Hz/60Hz (a):-15%/+10%

Electrical connections

terminals: 6-12mm , cable gland: 2xM20x1,5

Status alarm

1 free configurable status alarm with 2 potential free change-over
contacts, alarm point: < 2°C and > 8°C (adjusted at works)
contact rating 250VAC, 2A; 500 VA; 50W

Explosion proof protection

230V / 115V:
II 3 G EEx nAC [nL] IIC T4

02 ATEX 1040X

240V / 120V:
Class I, Div. 2, Groups A/B/C/D, T4
CAN/CSA-C22.2 No 61010.1-04 and No. 213-M87

120V:
Class I, Div. 2, Groups A/B/C/D, T4
FM Approval Standard Class No. 3611

Electrical safety

EN

EN 61010

CAN/CSA-C22.2 No.61010.1-04
UL Std. No. 61010-1(2 Edition)

Approval number

02 ATEX 1040X

System of protection

IP20 (EN 60529)

Case colour

RAL 9003

Method of mountingrt

wall mounting

Dimensions
Weight

270 x 270 x 316mm (BxHxT)
12kg/13,5kg (version: 230V/115V)