beautypg.com

Options for basic cooler ecm-exii, Description, Figure 2 – M&C TechGroup ECM-ExII Operator's manual User Manual

Page 10: Ecm-2g exii, 2 options for basic cooler ecm-exii, 11 description

background image

10

Gas sampling and gas conditioning technology

3-3.0EX-ME

10.2

OPTIONS FOR BASIC COOLER ECM-EXII

Heat exchanger type

ECM-1G

ECM-1PV

ECM-1SS

ECM-2G

ECM-2PV

ECM-2SS

Part No

93K0140

93K0170

93K0160

97K0100

97K0110

97K0115

Material of heat
exchanger

Borosilicate
glass

PVDF

st. steel
316Ti

Borosilicate
glass

PVDF

st. steel
316Ti

Max. gas flow rate

250Nl/h

250Nl/h

250Nl/h

150Nl/h

150Nl/h

150Nl/h

Sample gas pressure
abs.

max. 2bar
(3bar)

1

max. 3bar

max. 10bar

max. 2bar
(3bar)1

max. 3bar

max. 10bar

Sample gas connection

GL18
for Ø6mm
a.d. tube*

G1/4“i

G1/4“i
oder 1/4“
NPT**

GL18
for Ø6mm a.d.
tube*

Tube
Ø6mm*

Tube
Ø6mm*

Condensate connection

GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**

G3/8“i

G3/8“i o.
3/8“NPT**

GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**

G3/8“i

G3/8“i o.
3/8“NPT**

P at max. flow

1mbar

1mbar

1mbar

1mbar

1mbar

1mbar

Stagnant space approx.

100ml

100ml

100ml

40ml

25ml

30ml

Peristaltic pump SR25.2

1 pc. incorporated in the cooler, compl. installed, part no.: 01P9125; cooler weight
plus 0,6kg per pump

the sample gas pressure is limited to 2bar absolute

* Standard, other version on request

1

with GL connecting adapter

** Option
*** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew

point of 5 °C.



11

DESCRIPTION

Figure 2 shows the ECM-2G ExII cooler unit

.

Min. side distance

100

100

EC

M

Cooler

220

240

30

208

270

204

270

46

11

6

84

50

ventilation

opening

Compressor

ventilation outlet

Dimensions in mm

Figure 2 ECM-2G ExII