Options for basic cooler ecm-exii, Description, Figure 2 – M&C TechGroup ECM-ExII Operator's manual User Manual
Page 10: Ecm-2g exii, 2 options for basic cooler ecm-exii, 11 description
![background image](/manuals/255918/10/background.png)
10
Gas sampling and gas conditioning technology
3-3.0EX-ME
10.2
OPTIONS FOR BASIC COOLER ECM-EXII
Heat exchanger type
ECM-1G
ECM-1PV
ECM-1SS
ECM-2G
ECM-2PV
ECM-2SS
Part No
93K0140
93K0170
93K0160
97K0100
97K0110
97K0115
Material of heat
exchanger
Borosilicate
glass
PVDF
st. steel
316Ti
Borosilicate
glass
PVDF
st. steel
316Ti
Max. gas flow rate
250Nl/h
250Nl/h
250Nl/h
150Nl/h
150Nl/h
150Nl/h
Sample gas pressure
abs.
max. 2bar
(3bar)
1
max. 3bar
max. 10bar
max. 2bar
(3bar)1
max. 3bar
max. 10bar
Sample gas connection
GL18
for Ø6mm
a.d. tube*
G1/4“i
G1/4“i
oder 1/4“
NPT**
GL18
for Ø6mm a.d.
tube*
Tube
Ø6mm*
Tube
Ø6mm*
Condensate connection
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
G3/8“i
G3/8“i o.
3/8“NPT**
GL25 for
Ø12mm tube*
Ø8mm or
Ø10mm**
G3/8“i
G3/8“i o.
3/8“NPT**
P at max. flow
1mbar
1mbar
1mbar
1mbar
1mbar
1mbar
Stagnant space approx.
100ml
100ml
100ml
40ml
25ml
30ml
Peristaltic pump SR25.2
1 pc. incorporated in the cooler, compl. installed, part no.: 01P9125; cooler weight
plus 0,6kg per pump
the sample gas pressure is limited to 2bar absolute
* Standard, other version on request
1
with GL connecting adapter
** Option
*** Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew
point of 5 °C.
11
DESCRIPTION
Figure 2 shows the ECM-2G ExII cooler unit
.
Min. side distance
100
100
EC
M
Cooler
220
240
30
208
270
204
270
46
11
6
84
50
ventilation
opening
Compressor
ventilation outlet
Dimensions in mm
Figure 2 ECM-2G ExII