beautypg.com

Technical data, Receipt of goods and storage, Installation instructions – M&C TechGroup N9 KP18 Operator's manual User Manual

Page 9

background image

9

Gas sampling and conditioning technology

6-1.2.1ME

8

TECHNICAL DATA

Diaphragm Pump Series

N3 KPE / N3 KP18

N5 KPE / N5 KP18 N5 KPE / N5 KP18

Part Number

02P3351 / 02P3006 02P3355 / 02P3007 02P3360 / 02P3008

Max. operating pressure

0,25 bar

– 3 bar abs. 0,15 – 3 bar abs.

0,15

– 1,5 bar abs.

Capacity at atm. pressure

3 Nl/min

5 Nl/min

9 Nl/min

Ambient temperature

+5°C...+40°C

Sample temperature

+5°C...+40°C

Storage temperature

-15°C bis +60°C

Mains power supply

230V/50Hz; 115V/60Hz

Power consumption

60W

0,75A(230V)/1,5A(115V)

60W

0,65A(230V)/1,2A(115V)

Electrical connection

KPE: Single wire 4x0,5mm²
KP18: wired by luster terminal

Electrical protection

N3/5KPE: must be fuse-protected with 1 fuse: 230V-3,7A

T

/ 115V-

7,5A

T

N9KPE: must be fuse-protected with 1 fuse: 230V-3,15A

T

/ 115V-

6A

T

KP18: is fuse-protected with 2 fuses: 230V / 115V-3,15A

T

Preventing device

Thermal switch, double (115V/230V)

Connections

G 1/8“ i

DIN ISO 228/1

Protection

KPE: IP00 KP18: IP20

Housing material KP18

Al-Mg-Si-0,5 / firmness: F25

Material of sample contact-
ing parts

PVDF, Viton, PTFE

Weight

KPE : 1,1 kg
KP18 : 1,9 kg

KPE : 1,3 kg
KP18 : 2,1 kg



9

RECEIPT OF GOODS AND STORAGE

Please take the sample pump and possible special accessories carefully out of the packaging

immediately after receipt, and compare the goods with the items listed on the delivery note!

Check the goods for any damage caused during delivery and, if necessary, inform your transport

insurance company without delay of any damage discovered

N O T E !

The equipment should be stored in a protected, frost-free room!



10

INSTALLATION INSTRUCTIONS

All accident prevention regulations and safety instructions must be observed during installation and
operation. Please follow the instructions in chapter 3.