beautypg.com

Introduction, Application, Ambient conditions – M&C TechGroup N9 KP18 Operator's manual User Manual

Page 7

background image

7

Gas sampling and conditioning technology

6-1.2.1ME

6

INTRODUCTION

The compact diaphragm pumps type N3/5/9 are suitable for sampling air, gases and vapours in a
temperature range between + 5 °C and + 40 °C. The construction and the pump capacity are adapted
to the requirements of the gas analysis technique.



7

APPLICATION

The sampling evacuation and compression of the diaphragm sample pump series N3/5/9 is 100% oil-
free. The operation is gas-tight and maintenance-free.
The noise level of the pump without housing is less than 55 dB(A).(pneumatically connected).

Partnumber

Type

Material

Pump head

Diaphragm

Valve

Sealing

02P3351

N 3 KPE

PVDF

PTFE-beschichtet

Viton

Viton

02P3006

N 3 KP18

02P3355

N 5 KPE

02P3007

N 5 KP18

02P3360

N 9 KPE

02P3008

N 9 KP18

Material abbreviations according to ISO 1629 and 1043.1

Table 1

Pump materials

N O T E !

The diaphragm pumps series N 3/5/9 are not applicable for sam-
pling liquids.

Before use please check the compatibility of the material (see table
1).


7.1

AMBIENT CONDITIONS

When the pump is operating the following ambient conditions must be maintained:

Ambient temperature during operating: + 5°C .... + 40°C.

The pump must be protected from the effects of dust and water.

During operating an adequate supply of air for cooling must be provided.

W A R N I N G !

The pumps series N 3/5/9 are not allowed to be used in hazardous
areas.