Gas sampling and gas conditioning technology
2-0.1.1-ME
The M&C gas sample probes SP10 respectively SP10-H are used for continuous gas sampling in
processes with dust loadings up to 10g/m3, operating pressures up to max. 6 bar abs., temperatures
up to max. 600
C and high gas humidities. The modular design allows the combination of different
pre-filter materials (max. 900°C) and length (>10g/m
3
) and therefore an optimum adaption to the proc-
ess conditions. The compact design requires only little space.
N O T E !
Install the probe at a weather-protected sample point. For outdoor applica-
tions please use a version with option weather protection shield.
9
DESCRIPTION
The M&C gas sample probes are designed for easy installation, reliable operation and trouble-free
maintenance. Advantages are:
Gas sampling with dust loaded processes;
Low volume, fast response time;
Different pre-filters as an option.
The gas sample probe is heated up to max. +200
C.
10
PROBE DESIGN
The probe head with its all-round heating element forms a unit with the standard mounting flange
DN65 PN6 and the laterally mounted junction box.
A mounting clamp and a 1/8” internal thread are located on the underside of the probe for the connec-
tion of heated M&C gas sampling lines with external diameters of 40mm to a max. 55mm and a tube
diameter of max. 10mm.
After assembly of the sample line and the tube connector the sample gas outlet connection is en-
closed with the heat conducting shells. The maximum operating temperature for stainless steel pre-
filters type V10.. is 600
C and for the pre-filters V10..HC 900°C.
The modular system of our gas sample probes allows the usage of all M&C pre-filters type V10 and
extension tubes type Vo and Vm
with 1” thread. This guarantees an optimal adaptation to the process
conditions.