25 fixed value output – Toshiba LF516 User Manual
Page 94
93
6 F 8 A 0 9 3 4
●Changing the Mag-Prover’s self-diagnosis function setting
The following is an example to change the Mag-Prover’s self-diagnosis setting from OFF to ON.
Switch operation
Display example
Description
Select "SELF CHK" from the setting item
selection menu.
The current setting (OFF in this example)
appears.
Then push .
* Pushing returns you to the menu
screen.
The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)
Then push .
You can continue to change the setting item.
Push to change the selection
items.
: Selected item is scrolled up.
: Selected item is scrolled down.
When the desired item is selected, push
to set the item temporarily.
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
8.2.25 Fixed Value Output
The fixed value output function is used to output a fixed current and/or a fixed pulse output
independently of the flow rate signal. (The fixed pulse output is available only when Digital Output 1
(DO1) or Digital Output 2 (DO2) is used for pulse output function. For DO2, output can be obtained
only when fixed pulse output is 100pps or less.
The fixed-value output can be set in the ranges described below. (Current output and pulse output can
be set and output at the same time.)
• Fixed current output:
2.4 to 24 mA (can be set in increments of 0.1 mA)
• Fixed pulse output:
0 to 3000 pps (can be set in increments of 1 pps)
If fixed output is set to ON, Display 2 screen is used to indicate the fixed output in the measurement
mode.