Appendix 1 – Toshiba LF516 User Manual
Page 153
152
6 F 8 A 0 9 3 4
Appendix 1
Factory default standard value table
Parameter names
Default value(SI unit)
Default value(US unit)
Changed value
Excitation frequency
Value(*1)
Value(*1)
Flow direction
NORMAL
NORMAL
Display 1
m
3
/h gal/min
Display 2
m
3
B COUNT
B
Display digit setting
(for Display 1 and Display 2)
1/1000 1/1000
Custom coefficient
1.0
1.0
Custom unit
“ CUSTOM”
(Head of character string is
blank )
“ CUSTOM”
(Head of character string is
blank )
Range type
SINGLE SINGLE
Range 1
Value(*1)
Value(*1)
Ranges 2 to 4
0.00 m
3
/h 0.00
gal/min
Hysteresis
3.0 %
3.0 %
Damping constant
5.0 s
5.0 s
Rate-of-change limit
0.0 %
0.0 %
Control limit time
0.0 s
0.0 s
Low cutoff
1.0 %
1.0 %
Display low cutoff
LINEAR
LINEAR
Manual zero
0.0 %
0.0 %
4
−20mADC alarm output
4mA 4mA
Output low limit setting
4mA
4mA
Digital output 1
NO USE
NO USE
Digital output 2
NO USE
NO USE
DO1/DO2 active status
NormOPEN
NormOPEN
Digital input
NO USE
NO USE
DI detective level
H LEVEL
H LEVEL
Count rate
Value(*1)
Value(*1)
Pulse width setting mode
AUTO
AUTO
Pulse width
100 ms
5 ms
Preset count
00000000
00000000
Preset function
HOLD
HOLD
High alarm On/Off
OFF
OFF
High alarm value
0.0 %
0.0 %
Low alarm On/Off
OFF
OFF
Low alarm value
0.0 %
0.0 %
High-High alarm On/Off
OFF
OFF
High-High alarm value
0.0 %
0.0 %
Low-Low alarm On/Off
OFF
OFF
Low-Low alarm value
0.0 %
0.0 %
Self-diagnosis On/Off
ON
ON
Fixed value output
OFF
OFF
Fixed value current
4mA
4mA
Fixed value pulse
0 pps
0 pps
Password 000 000
LCD density adjustment
3
3
Switch position setting
BOTTOM
BOTTOM
*1 : Setting value by meter size please refer to the next list.