1 0 % flow rate calibration (zero calibration) – Toshiba LF516 User Manual
Page 105
104
6 F 8 A 0 9 3 4
9.2 Calibration Using Mag-Prover’s Built-In Signal Source
9.2.1 0 % Flow Rate Calibration (Zero Calibration)
Using Mag-Prover’s internal calibration circuit, 0% flow rate calibration (hereafter called zero
calibration) can be performed.
●Zero point check / calibration
Switch operation
Display example
Description
Select "CAL 0" from the setting item
selection menu.
When the calibration screen is selected, the
internal simulation circuit starts working and
0%
value using the internal simulation
signal appears.
Then push and hold longer.
* Pushing returns you to the setting
menu.
The title of the screen changes to “ADJ
READY
” and the converter is is ready for
calibration.
Pushing returns you to the previous
screen.
Push to start calibration for 0% flow
rate.
It takes several seconds to perform
calibration for 0% flow rate and the
simulated value of 0% after calibration
appears.
Pushing returns you to the setting
menu.
Note 1: To perform calibration, push and hold longer.
Note 2: To cancel the adjustment when ADJ READY is displayed, push .
The screen returns to the zero display using the simulation input.