LF546 Converter Specifications
Input
Input signal:
Flow rate proportional signal from the detector
Digital input signal
Signal type:
20 to 30Vdc voltage signal
Input resistance: About 2.7k
Number of input points:
1
Digital input function: Select either of the following.
Range switching input: Large/Small range switching of unidirectional double range,
forward/reverse direction double range
Counter control input: Internal totalization counter start/stop/reset control
Output hold input: The current output and pulse output are kept to their preset
values.
Zero adjustment input: Start still water zero adjustment.
Output
Current output:
4 to 20mAdc (load resistance 750 or less)
Digital output 1:
Output type:
Transistor open collector
Capacity:
30Vdc, Max 200mA
Number of output points:
1
Digital output 2:
Output type:
Semiconductor contact signal
output
(no polarity)
Capacity:
150Vdc, Max. 150mA
150Vac (peak value), Max.
100mA
Number of output points :
1