24 mag-prover, Self diagnosis on/off setting – Toshiba LF516 User Manual
Page 93
92
6 F 8 A 0 9 3 4
Switch operation
Display example
Description
Push to move the cursor to the
digit you want to change and push
to change the number of the digit.
When you push , a message appears
to confirm whether the setting is OK or not.
If OK, push . If you want to redo the
setting, push .
Pushing cancels the setting operation
and exits the setting screen.
Note: If you try to set a value outside of the range
−10% to +110%, LOW OVER SPEC or
HIGH OVER SPEC
error appears and the value returns to the previous value. Set a
value once again.
8.2.24 Mag-Prover
-Self Diagnosis ON/OFF Setting
You can select on/off setting for Mag-Prover’s self-diagnosis function.
If the self-diagnosis function is set to OFF, no error message is displayed even if any of the errors listed
below occurs.
• ROM error
• RAM error
• System parameter error
• Excitation cable is not connected or its wiring is open
• Excitation circuit failure
• ADC circuit error
• Totalizer data is destroyed
Selection items
Description
OFF
Self-diagnosis function is turned off.
ON Self-diagnosis
function is turned on.
If this function is set to ON and an error occurs, an error message appears on the Display 2 measured
value screen. If an error occurs, the measurement items specified for Display 2 screen cannot be
displayed unless the error is removed.