Toshiba LF516 User Manual
Page 130
129
6 F 8 A 0 9 3 4
(2) Span setting
Setting of 8.2.7 “Custom Coefficient Setting” and 8.2.8 “Custom Unit Setting” is applied to the
custom coefficient and unit same as (1)Display of flow rate.
Example : In the case of custom unit [dL(deciliter)/min].
Custom coefficient setting
Set the custom coefficient in accordance with 8.2.7 “Custom Coefficient
Setting”. Set the conversion coefficient from m
3
/min
unit to the custom
coefficient. In the case of dL/min, set 10000 that is the conversion coefficient
of dL from m
3
.
Custom unit setting
Set the custom unit in accordance with 8.2.8 “Custom Unit Setting”. Set the
character string of unit to the custom unit. In the case of this example, set ‘d’
‘L’ ‘/’ ‘m’ ‘i’ ‘n’. The rest of 1 character is blank space because the maximum
character number of the custom unit is 7.
Span
setting
Select the custom unit dL/min to the span unit in accordance with 8.2.9 “Span
(Range)”. When custom unit is selected, the new span value will be displayed
automatically based on the custom coefficient and custom unit.