Toshiba LF516 User Manual
Page 82
81
6 F 8 A 0 9 3 4
●Digital output functions
Selection items
Digital output functions
NO USE
Not used
H ALM
High alarm output
L ALM
Low alarm output
HH ALM
High-High alarm output
LL ALM
Low-Low alarm output
RNG SIG1
Range output No. 1
RNG SIG2
Range output No. 2
PRESET C
Preset count output
CONV ALM
Converter failure alarm output
PLS OUT
Pulse output
PLS FRD
Fixed forward flow pulse output
PLS REV
Fixed reverse flow pulse output
MRH ALM
Multi-range high alarm output (option)
MRL ALM
Multi-range low alarm output (option)
Notes: When the range type is set to Forward/reverse multiple ranges, and if the pulse output
(PLS OUT) is selected, pulses of forward and reverse directions will be output.
For setting method of the range type, see 8.2.9, “Span (Range).”
●Digital output active status (Only when alarm output is set )
Selection items
Alarm output action
NormCLOSE
Normal: Contact closed, Alarm out: Contact open
NormOPEN
Normal: Contact open, Alarm out: Contact closed
●Changing the digital output function
The following is an example to change the Digital Output 1 (DO1) function from High alarm
output (H ALM) to Low alarm output (L ALM).
Switch operation
Display example
Description
Select "DO1 FUNC" from the setting item
selection menu.
The current setting (H ALM in this example)
appears.
Then push .
* Pushing returns you to the menu
screen.