6 F 8 A 0 9 3 4

● For pulse width setting mode, you can select either AUTO or MANUAL.

Depending on this setting, the pulse width setting varies as shown in the table below:

Selection item

Pulse width value to be set

AUTO

After the count rate is set, the pulse width is automatically set
to 40% of the period of pulse frequency at 100% output.

MANUAL

Even after the count rate is set, the pulse width is not changed.

However, if the pulse width becomes out of the setting

range as a result of count rate setting, the screen is
automatically switched to the pulse width setting screen
after the count rate is set.

Note: If the count rate exceeds 1000 (pulse/s), the pulse width setting mode is limited to the AUTO

mode only and you cannot set the width manually.

● Examples of pulse width setting range

Example 1
In the case the range is 108m

3

/h (0.03m

3

/s) and the count rate is 0.00003m

3

:

Since the pulse rate is 108 (m

3

/h) / 0.00003(m

3

) = 3600000 pulse/h (1000 pulse/s), the period

of pulse frequency at full scale is 1ms.
Therefore, the pulse width can be set only to: 1ms

× 40% = 0.4ms only.

Example 2
In the case the range is 108m

3

/h (0.03m

3

/s) and the count rate is 30m

3

:

Since the pulse rate is 108(m

3

/h) / 30(m3) = 3.6 pulse/h (0.001 pulse/s), the

period of pulse frequency at full scale is 1000000ms.
Therefore, the pulse width is: 1000000ms

× 40% = 400000ms. However, since the

maximum value is 500ms, the pulse width becomes 500ms.

Example 3
In the case the range is 108m

3

/h (0.03m

3

/s), the count rate is 0.03m

3

and the pulse width is set

to 0ms:

Since the pulse rate is 108(m

3

/h)

× 0.03(m

3

) = 3600 pulse/h (1 pulse/s), the period

of pulse frequency at full scale is 1000ms.
Therefore, the pulse width is: 1000ms

× 40% = 400ms. However, since the

maximum value is 100ms in the case of Auto setting, the pulse width
becomes 100ms.