6 F 8 A 0 9 3 4

5.4.2 Cable Connection


Connect and install the terminal-treated cables to the terminal block by the following procedure.

Connect the cables to the terminal block securely. A loose connection may cause incorrect

measurement. After connecting a cable, try to pull it to check whether it has been connected
securely.

Referring to Section 5.2 "External Device Connections and Grounding", connect each cable to the
terminal block. Tighten the screws of the terminal block tightly to ensure the secure connection. A
loose connection may cause incorrect measurement. After connecting a cable, try to pull it to see
whether it has been connected securely.

Figure 5.10 Connecting a Cable to Terminal Block

Phillips screwdriver

Terminal block

The appropriate torque for tightening

the terminal board screw is 1.2 N

m.