●Digital input control signal
You can select the detective level of the digital input, as shown below, to control the totalizer
and pulse output.
(Only when the digital input function is set for totalizer control input)
Selection items
Digital input function setting
Totalizer control signal
CNT ST/SP
Totalizer START/STOP
H signal
Totalizer STOP
L signal
Totalizer START
L LEVEL
CNT RS/ST
Totalizer RESET/START
H signal
Totalizer START
L signal
Totalizer RESET
CNT ST/SP
Totalizer START/STOP
H signal
Totalizer START
L signal
Totalizer STOP
H LEVEL
CNT RS/ST
Totalizer RESET/START
H signal
Totalizer RESET
L signal
Totalizer START
●Changing the digital input function
The following is an example to change the Digital Input (DI) function from No use (NO
USE
) to Totalizer Start / Stop (CNT ST/SP).
Switch operation
Display example
Description
Select "DI FUNC" from the setting item
selection menu.
The current setting (NO USE in this
example) appears.
Then push .
* Pushing returns you to the menu
screen.
The switches at the bottom change.
( are shown.)
At the same time, the cursor appears.
(The item indicated by the cursor is
highlighted.)
Then push .
You can continue to change the setting item.
Push to change the selection
items.
: Selected item is scrolled up.
: Selected item is scrolled down.
When the desired item is selected, push
to set the item temporarily.