background image

Retrieving Test Results - 7

Page 50

14439.C 03/2013

VerifyNow System User Manual |

USA

7.4

View Event Log

The instrument stores the last 50 error/attention events and the last 11 alarms in its memory. The

instrument Event log displays the errors and alarms that have occurred during use. They are arranged

chronologically. Use the following procedure to view the Event log (Figure 7-7):

1.

From the main menu, enter

Operator ID and

Password, if required. Use the keypad to enter

the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to

enter the password number.

2.

Press the Maintenance

key.

3.

Press the Next

key two times. Refer to the Appendix for information on navigating within the

Maintenance section.

4.

Press the Event Log

key to retrieve the appropriate usage information.

See Section 7.4.1 for a description of the event log.

7.4.1

Alarm and Error/Attention Log

The cursor is positioned on the most recent event (e.g. errors, alarms, alerts). Navigate up and down an

individual list using the ( , ) keys or between lists (Alarms, Errors) using the Tab key.

26-Feb-2012

14:57:00

CODE OpId MMDDHHMM

021IL 6789 06061230

24W 7779807042359

36A 9876508110900

090 333 20090915

ALL

Alarms

Errors

Figure 7‑7 Alarm and Error/Attention Log

Error/Attention List

Alarms List

Print Result

Print All Results

Back

Feature

Description

Alarms List

Lists the most recent of alarms logged during operation. This includes the alarm

code, operator identification (if enabled), and the date and time stamp in month, day,

hour (24 hr clock format), and minute, in “MMDDHHMM” format.

Error/Attention

List

Lists the most recent errors or alerts (attentions) logged during operation. This

includes the error/alert code, operator identification (if enabled), and the date and

time stamp in month, day, hour (24 hr clock format), and minute, in “MMDDHHMM”

format.

In addition to viewing earlier results, icons keys enable results to be printed (optional) or another test

started.