Page 50
Retrieving Test Results - 7
Page 46
14439.C 03/2013
VerifyNow System User Manual |
USA
In addition to viewing earlier results, icons keys enable results to be printed (optional) or another test
started.
Icon
Description
New Test
Select this icon to perform another test on the same patient
(when Patient ID is enabled). The instrument will display the
Patient ID screen with the Patient ID complete
d.
Print Result
Scroll to the selected record, and select this icon to print the
selected test result
s.
Print All
Results
Scroll backwards through results using the ( ) key, and
select this icon to print all test results in the log that are
positioned forward of the cursor, or all of the test results for a
particular patient (when Patient ID function is activat
ed).
Back
Select this icon to return to the Main Menu
.
7.2
Retrieving Quality Control
The instrument stores the last 100 EQC and WQC results in its memory. You can retrieve and print
information about the earlier EQC parameters, WQC results, test type, date and time, and operator
identification (when configured). Use the following procedure to view the quality control results.
1.
From the main menu, enter
Operator ID and
Password, if required. Use the keypad to enter
the Operator ID number at the prompt. Press the Tab key to advance to the Password prompt to
enter the password number.
2.
Press the Maintenance
key.
3.
Press the Next
key three times. Refer to the Appendix for information on navigating within the
Maintenance section.
4.
Press the EQC
key or the WQC
key to retrieve the appropriate results.
See Section 7.2.1 for a description of the EQC results log or Section 7.2.2 for a description of the WQC
results log.
7.2.1
Electronic Quality Control Results Log
The cursor is positioned on the most recent test result (Figure 7-4). Navigate within an individual EQC
test result using the ( , ) keys.
26-Feb-2012 13:51:12
EQC Results
P/F 01 02
Op V1 V2
MMDD P1 F1
HHMM F2 Em
P
0.30 1.20
00005 5.45 0.25
0429 14 14
1327 0000
F
0.30 1.20
00005 3.45 0.23
0429 0.35 14
1241 14 F080<-
S/N 1095
ALL
Figure 7‑4 EQC Results Log
EQC Result
Operator ID
Date and Time
Parameters
Flagged Parameter
Print Result
Back
Print All Results