ITC VerifyNow System Platelet Reactivity Test User Manual
Page 51

Page 47
14439.C 03/2013
Retrieving Test Results - 7
VerifyNow System User Manual |
USA
Feature
Description
EQC Result
Displays the result as either a Pass (P) or Fail (F). A Pass result
indicates the instrument was operating within its specification limits.
If there was an earlier problem detected during the EQC process,
then a FAIL result would display.
Operator ID
Displays the operator identification for the user who performed the
test (when configured).
Date and Time
Stamp
Displays the date and time by month, day, hour (24 hr clock
format), and minute using the following notation, “MMDD HHMM.”
Parameter
Displays the result for the following measured parameters:
O1: High optical signal level
O2: Low optical signal level
V1: Maximum vacuum level
V2: Maximum leak rate
P1: Maximum pressure level
F1: Mixer frequency flag count
F2: Mixer frequency transitions high-to-low
Mask: Identifies failing component element
Flagged
Parameter
The failed parameter(s) will be indicated by an arrow ( ) to
the right of the measured value(s). Refer to the Appendix for a
description of parameters.
In addition to viewing earlier results, icons keys enable results to be printed (optional).
Icon
Description
Print Result
Scroll to the selected record, and select this icon to print the
selected EQC resul
ts.
Print All
Results
Scroll backwards through the results using the ( ) key, and
select this icon to print all EQC results in the log that are
forward of the cursor posit
ion.
Back
Select this icon to return to the previous screen
.