Verifynow, System – ITC VerifyNow System Platelet Reactivity Test User Manual
Page 30

Quality Controls - 5
Page 26
14439.C 03/2013
VerifyNow System User Manual |
USA
QC
26-Feb-2012
14:57:00 104A
VerifyNow
System
®
26-Feb-2012
14:57:00
Figure 5‑19 Start WQC Level 2
5.
Open the foil pouch and remove the test device just before use. Hold it by the finger grip.
NOTE: Each test device has a finger grip. Avoid handling other surfaces of the test device.
6.
Remove the needle’s protective sheath by pulling directly up on the sheath. Do not twist the sheath,
as this may remove the needle (Figure 5-20).
Figure 5‑20 Prepare Needle
7.
Open the cover. Using the finger grip, insert the test device into the test device port until it clicks
(Figure 5-21).
Figure 5‑21 Insert Device
8.
If the test device is the first from a new lot, the instrument will display a bar code screen as soon as
the test device is inserted into the port and the device spot code is read. (Figure 5-22). Position the
barcode of the test device pouch in front of the bar code scanner on the left side of the instrument
so that the bar code on the bottom edge of the pouch lines up with the scanner window. Move the
pouch both towards and away from the red barcode light. An audible beep will be heard when the
instrument reads the information. Press the Retry
key if you are unable to scan the bar code the
first time.