6.3

Sample Handling

Fresh whole blood samples in the appropriate partial-fill collection tubes are required for use with

the VerifyNow instrument. Keep samples at room temperature. Do not centrifuge, separate, freeze or

refrigerate sample. Table 6-1 provides the amount of time blood must incubate at room temperature

after collection and before test. Blood samples older than indicated must be discarded and a new

sample drawn.

Table 6‑1 Test Times

Test

Wait Time

Test Window

Aspirin

30 min

30 min to 4 hrs

PRUTest

10 min

10 min to 4 hrs

IIb/IIIa

None

Within 15 minutes

Samples should be handled according to your institution’s policies and procedures pertaining to

biohazardous materials.

6.4

Performing a Test

Materials

You will need the

1 Test device and 1 Sample tube:

NOTE: Test device should remain sealed in the foil pouch until ready for use to prevent damage by humidity.

NOTE: Do not use test devices or WQC materials beyond the expiration date.

Procedure

1.

Locate one test device and the sample tube. If test device is refrigerated, allow it to reach room

temperature (18-25ºC or 64 -77ºF) prior to use. Do not remove the test device from the foil pouch

during this step.

2.

Power on the instrument using the power switch on the back panel (“I” designates on). The

instrument will power on and perform a self-testing routine lasting approximately 30 seconds. After

the self-testing is complete, the Start screen will display. Press the Next key to advance to the Main

Menu (Figure 6-2). The instrument should be allowed to warm-up for at least 15 minutes prior to use.

26-Feb-2012

14:57:00 104A

444

QC

VerifyNow

System

®

The System is

warming up.

Figure 6‑2 Main Menu

A flashing message will display while the instrument is initially warming up. During this time the icon

keys are inactive and you will be unable to start a test. You will not be able to go to the next step

until after the warm up message no longer displays.