beautypg.com

ITC VerifyNow System Platelet Reactivity Test User Manual

Page 27

background image

Page 23

14439.C 03/2013

Quality Controls - 5

VerifyNow System User Manual |

USA

10.

Wait for the image of the tube to display, then insert the WQC sample into the sample well of the

test device with the rubber stopper facing downward, so that the needle fully pierces the stopper.

The instrument will produce two audible beeps when the tube has been fully inserted (Figure 5-12).

26-Feb-2012

WQC P2Y12

14:57:00

Figure 5‑12 Insert Tube

11.

Close the cover to the test device port (Figure 5-13).

Figure 5‑13 Close Cover

12.

The instrument automatically draws the sample from the vacuum collection tube into the test device,

heats the sample, and proceeds with the analysis of the sample. During this time the screen will

flash the test device icon, indicating that the sample is processing. Do not open the cover until the

test has completed and a result displays. A calculator will display when the test is near completion

(Figure 5-14).

Figure 5‑14 Test Progress

WQC P2Y12

26-Feb-2012

14:57:00

KEEP COVER CLOSED.

999

WQC P2Y12

26-Feb-2012

123

14:57:00

KEEP COVER CLOSED.

WQC P2Y12

26-Feb-2012

123

14:57:00

CAUTION:

The sample is pressurized during parts of the test. Never remove the sample

tube or test device during the test. To abort a test, press the Back

key. Wait for the

prompt before removing the test device and tube.

P2Y12