Quality Controls - 5

VerifyNow System User Manual |

USA

Figure 5‑22 Scan Bar Code

Bar Code

WQC P2Y12

U s e b a r c o d e

s c a n n e r t o r e a d

t h e b a r c o d e .

26-Feb-2012

14:57:00

After the bar code has been scanned once, the instrument will accept all remaining test devices from

that lot without displaying the bar code screen.

NOTE: If you are experiencing difficulty with the bar code scanner scanning the label, try

adjusting the angle of the scan by raising the corner of the instrument.

9.

Inspect the pellet. It should appear pink. If the pellet appears smaller, red or is stuck to the vial,

discard and use a new pellet (Figure 5-23).

Figure 5‑23 Inspect Pellet

10.

Immediately before use, open the vial containing the pellet. Remove the cap from the diluent tube by

twisting and pulling the cap simultaneously.

11.

Hold the diluent tube vertically. Then, invert the vial containing the pellet so that the pellet falls into

the diluent tube (Figure 5-24). Ensure the pellet does not stick to the side of the diluent tube, as it

will not reconstitute properly. If you are unsure, discard the tube and retry using another WQC

diluent tube and pellet.

Figure 5‑24 Add Pellet