Quality Controls - 5
VerifyNow System User Manual |
USA
Figure 5‑22 Scan Bar Code
Bar Code
WQC P2Y12
U s e b a r c o d e
s c a n n e r t o r e a d
t h e b a r c o d e .
26-Feb-2012
14:57:00
After the bar code has been scanned once, the instrument will accept all remaining test devices from
that lot without displaying the bar code screen.
NOTE: If you are experiencing difficulty with the bar code scanner scanning the label, try
adjusting the angle of the scan by raising the corner of the instrument.
9.
Inspect the pellet. It should appear pink. If the pellet appears smaller, red or is stuck to the vial,
discard and use a new pellet (Figure 5-23).
Figure 5‑23 Inspect Pellet
10.
Immediately before use, open the vial containing the pellet. Remove the cap from the diluent tube by
twisting and pulling the cap simultaneously.
11.
Hold the diluent tube vertically. Then, invert the vial containing the pellet so that the pellet falls into
the diluent tube (Figure 5-24). Ensure the pellet does not stick to the side of the diluent tube, as it
will not reconstitute properly. If you are unsure, discard the tube and retry using another WQC
diluent tube and pellet.
Figure 5‑24 Add Pellet