ITC VerifyNow System Platelet Reactivity Test User Manual
Page 33
Page 29
14439.C 03/2013
Quality Controls - 5
VerifyNow System User Manual |
USA
Figure 5‑28 Test Progress
WQC P2Y12
26-Feb-2012
14:57:00
KEEP COVER CLOSED.
999
WQC P2Y12
26-Feb-2012
123
14:57:00
KEEP COVER CLOSED.
WQC P2Y12
26-Feb-2012
123
14:57:00
CAUTION:
The sample is pressurized during parts of the test. Never remove the
sample tube or test device during the test. To abort a test, press the Back
key.
Wait for the prompt before removing the test device and tube.
16.
When the results display, record the results or print
the results, if your instrument has an
attached printer (Figure 5-29).
NOTE: Re-draw sample if there is evidence of hemolysis or clotting in any other tube that was collected at the same time.
NOTE: Re-draw sample if there is evidence of hemolysis or clotting in any other tube that was collected at the same time.
26-Feb-2012
14:57:00
WQC P2Y12
PRU:
4
Figure 5‑29 WQC Level 2 Result
17.
Open the cover. Remove the test device and tube together in one piece by grasping the test device
finger grip and pulling straight up (Figure 5-30). Do not separate the tube from the test device. Close
the cover. Discard the test device and tube in a biohazard waste in accordance with institutional
guidelines.
Figure 5‑30 Remove Device