4 creating psd calibrationreference(.ref) files, Creating psd calibration reference (.ref) files, 4 creating psd calibration reference (.ref) files – HP Data Explorer 4 Series User Manual
Page 359
Calibrating a PSD Spectrum
Data Explorer
™
Software User’s Guide
8-21
8
8.3.4 Creating PSD
Calibration Reference (.REF) Files
Overview
You can manually create a calibration reference file by typing
masses in a text file as described in Section 5.3.3, Creating or
Modifying a Calibration Reference File (.REF).
You can also use the Ion Fragmentation calculator to generate
theoretical fragments and masses from the sequence for a
standard compound, then automatically save the masses and
associated information in a calibration reference file. For
detailed information on using the Ion Fragmentation calculator,
see Section 6.5, Using the Ion Fragmentation Calculator.
Using the Ion
Fragmentation
calculator to
create a
calibration
reference (.REF)
file
To create a PSD calibration reference file using the Ion
Fragmentation calculator:
1.
From the Applications menu, select Ion Fragmentation
Calculator.
The Ion Fragmentation Calculator dialog box is displayed
(Figure 8-4 on page 8-8).
2.
In the Sequence text box, type the amino acid sequence
of the standard compound. Use single-letter codes. Set
other parameters as needed. For parameter descriptions,
see Section 6.5, Using the Ion Fragmentation Calculator.
3.
Select Monoisotopic from the Calculate fragment ion
masses as drop-down list.