2 the resolution-based peak detection routine, The resolution-based peak detection routine – HP Data Explorer 4 Series User Manual
Page 101
Overview
Data Explorer
™
Software User’s Guide
3-3
3
3.1.2 The Resolution-Based
Peak Detection Routine
This section describes:
•
Type of data affected
•
Process that occurs
•
Detection ranges
•
PSD peak detection for Voyager data
Type of data
affected
The resolution-based peak detection routine applies to
spectral data only. You can enable or disable the
resolution-based peak detection routine as described on
page 3-14.
Process that
occurs
When enabled, the resolution-based peak detection routine
automatically:
•
Divides a trace into detection ranges based on the
expected number of data points across a typical mass
spectral peak.
•
Applies a Filter Width that is equal to the number of data
points per peak in each detection range.
•
Uses a Filter Increment of 1.