beautypg.com

Manufacturer id register, Device type register, Status/control register – HP E1459A User Manual

Page 79

background image

78

HP E1459A Register Definitions

Manufacturer ID

Register

ID register (base = 00

h

) is a read only register. For the Isolated Digital

Input/Interrupt, a read of the ID register returns FFFF

h

since the

multiplexers are manufactured by Hewlett-Packard and are A16 only,
register-based devices.

Manufacturer ID Register (base + 00

h

)

* Returns FFFF

h

= Hewlett-Packard A16 (only) register-based device

Device Type

Register

Device Type register (base = 02

h

) is a read only register. For the Isolated

Digital Input/Interrupt, a read of the Device Type register returns 0154

h

.

This indicates it is a model HP E1459A.

Device Type Register (base + 02

h

)

Status/Control

Register

Status/Control register (base = 04

h

) can be read and written. Many of the

bits perform control functions. Reading this register returns the current state
of the status bits for the module.

Status/Control Register (base + 04

h

)

NOTE: Bits 8 and 9 are returned in the IACK response in the same bit
positions.

WRITE
R
= Reset to power-on state by writing a "1" in this bit (Must be set back to
"0").
BS = Bank Select. When "0" Port 0 and Port 1 data are accessed in registers
b + 10

h

through b + 2E

h

. When "1" Port 2 and Port 3 data are accessed in the

same registers.
I = Interrupt Enable. When set to 1, an IRQ can be generated with an edge
event (assuming one is enabled).
D = Data Ready Enable. When set to 1 an IRQ can be generated with a
DAVX line is asserted.

b + 0

h

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Write

No Effect

Read

Manufacturer ID

b + 2

h

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Write

No Effect

Read

Always Returns 0154

h

b + 4

h

15

14

13

12

11

10

9

8

7

6

5

4

3

2

1

0

Write

Undefined

D

I

BS

Undefined

R

Read

Undefined

M

Undefined

D IRQ

E IRQ

Undefined

D

I

BS

Undefined

R