HP E1459A SCPI Command Reference

41

DIAGnostic:SYSReset Subsystem

The DIAGnostic:SYSReset Subsystem controls and monitors the Watchdog Timer.
Refer to Chapter 1 for detailed information on the Watchdog Timer.

Syntax

DIAGnostic:SYSReset

[:STATe]?

page 41

:ENABle <state>

page 41

:ENABle?

page 42

DIAGnostic:SYSReset[:STATe]?

Returns the value of the Watchdog Timer state from the Watchdog Timer Control /
Status Register.

Parameters

None

Comments

Returns a 1 if the Watchdog Timer is asserted; returns a 0 if the Timer is not

asserted.

Reading this register resets the Watchdog Timer.

If the Watchdog timer is enabled, the state of the timer MUST be read before

the Watchdog time elapses. If the state is not read before the time elapses, the
Watchdog Timer asserts the VXIbus SYSRESET line.

DIAGnostic:SYSReset:ENABle <

state>

Turns the Watchdog Timer ON or OFF.

Parameters

Comments

A 0 or OFF turns the Watchdog Timer off; a 1 or ON turns the Timer on.

CAUTION: When the Watchdog Timer is enabled (ON), the VXIbus

backplane SYSRESET line is asserted if the Watchdog Timer is allowed to
elapse. The Watchdog Timer is reset each time the state of the Timer is read by
the DIAG:SYSR:STAT? command.

Example

DIAG:SYSR:ENAB ON

Turns Watchdog Timer on

Parameter Name

Parameter Type

Range of Values

Default

<

state>

numeric or discrete

0, 1, OFF, ON

0, OFF