Precision Digital PD6100 User Manual
Page 9

Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual
9
ISOLATED
SENSOR POWER
SUPPLY
Terminals P+ & P-: 10 VDC or 5 VDC
10%.12/24 VDC
powered models selectable for 24 (should not be used for strain
gauge/load cell), 10, or 5 VDC supply (internal jumper, see
Sensor Excitation Voltage Selection on page 18).
ProVu Series (PD6100):
85-265 VAC models rated @ 200 mA max, 12/24 VDC powered
models rated @ 350 mA max for 5 or 10 VDC supply, @ 100 mA
max for 24 VDC supply.
ProtEX-MAX (PD8 Series):
All models transmitter supply rated @ 25mA max.
Note: Do not use 24 VDC to power strain gauge bridge
NORMAL MODE
REJECTION
Greater than 60 dB at 50/60 Hz
ISOLATION
4 kV input-to-power line
500 V input-to-output (powered by external supply)
OVERVOLTAGE
CATEGORY
Installation Overvoltage Category II:
Local level with smaller transient overvoltages than
Installation Overvoltage Category III.
ENVIRONMENTAL
ProVu Series (PD6100):
Operating temperature range: -40 to 65°C
Storage temperature range: -40 to 85°C
Relative humidity: 0 to 90% non-condensing
ProtEX-MAX (PD8 Series):
T6 Class operating temperature range Ta = -40 to 60°C
T5 Class operating temperature range Ta = -40 to 65°C
See LIM8 ProtEX-MAX instruction manual for additional
details.
MAX POWER
DISSIPATION
ProtEX-MAX (PD8 Series): Maximum power dissipation
limited to 15.1 W. See LIM8 ProtEX-MAX instruction
manual for additional details.
CONNECTIONS
Removable screw terminal blocks accept 12 to 22 AWG
wire, RJ45 for external relays, digital I/O, and serial
communication adapters.
ENCLOSURE
1/8 DIN, high impact plastic, UL 94V-0, color: black
MOUNTING
1/8 DIN panel cutout required:
3.622" x 1.772" (92 mm x 45 mm)
Two panel mounting bracket assemblies are provided.
TIGHTENING
TORQUE
Screw terminal connectors: 5 lb-in (0.56 Nm)
OVERALL
DIMENSIONS
4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm)
(W x H x D)