Maximum/Minimum Readings
The max & min readings (peak & valley) reached by the process can be
displayed either continuously or momentary:
1. Display briefly by assigning to the F1-F3 function keys or to the
digital inputs in the User menu. Any of the F1-F3 function keys
(buttons) and the digital inputs can be programmed to reset the
max & min readings.
2. Display continuously by assigning either display to max/min through
the Display menu.
To display max reading using function key:
1. Press user assigned function key/digital input to display maximum
reading since the last reset/power-up.
2. To reset max/min, press the Menu button, then the Up arrow/F2
button until the Reset (Reset) menu is displayed.
3. Press the Up arrow/F2 button to select reset high (rst Hi), reset
low (rst Lo), or reset high and low (rst HL) is displayed and then
press the Enter/F3 button. The max and/or min displays are reset to
actual values.
4. Press Menu to exit max/min display reading.
To display max/min readings continuously:
Assign either display to Max (d Hi), Min (d Lo), or toggle between Max
and Min (d HL) every 10 seconds.