Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual
73
Display Parameter Action/Setting
O-rang
Overrange
Program mA output for display overrange
u-rang
Underrange
Program mA output for display
underrange
mAx
Maximum
Program maximum mA output allowed
min
Minimum
Program minimum mA output allowed
Calib
Calibrate
Calibrate 4-20 mA output (internal
reference source used for scaling the
output)
4 mA
4 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
20 mA
20 mA output
Enter mA output value read by milliamp
meter with at least 0.001 mA resolution
user
User I/O
Assign function keys and digital I/O
F1*
F1 function key
Assign F1 function key (*F1/F2/F3)
F4
F4 function
Assign F4 function (digital input)
dI I
Digital input 1
Assign digital input 1 – 8, if expansion
modules are connected
dO 1
Digital output 1
Assign digital output 1 – 8, if expansion
modules are connected
ICal
Internal source
calibration
Enter internal source calibration (used for
scaling the meter without a signal source)
15mv
15 mV
calibration
Calibrate 15 mV input range (internal
reference source used for scaling the
input)
20mv
20 mV
calibration
Calibrate 30 mV input range (internal
reference source used for scaling the
input)
100mv
100 mV
calibration
Calibrate 150 mV input range (internal
reference source used for scaling the
input)
200mv
200 mV
calibration
Calibrate 300 mV input range (internal
reference source used for scaling the
input)
stroff
Strain offset
Calibrate the offset of the input circuit
v Lo
mV low
Calibrate low mV input (e.g. 0 mV)