Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual

73

Display Parameter Action/Setting

O-rang

Overrange

Program mA output for display overrange

u-rang

Underrange

Program mA output for display
underrange

mAx

Maximum

Program maximum mA output allowed

min

Minimum

Program minimum mA output allowed

Calib

Calibrate

Calibrate 4-20 mA output (internal
reference source used for scaling the
output)

4 mA

4 mA output

Enter mA output value read by milliamp
meter with at least 0.001 mA resolution

20 mA

20 mA output

Enter mA output value read by milliamp
meter with at least 0.001 mA resolution

user

User I/O

Assign function keys and digital I/O

F1*

F1 function key

Assign F1 function key (*F1/F2/F3)

F4

F4 function

Assign F4 function (digital input)

dI I

Digital input 1

Assign digital input 1 – 8, if expansion
modules are connected

dO 1

Digital output 1

Assign digital output 1 – 8, if expansion
modules are connected

ICal

Internal source
calibration

Enter internal source calibration (used for
scaling the meter without a signal source)

15mv

15 mV
calibration

Calibrate 15 mV input range (internal
reference source used for scaling the
input)

20mv

20 mV
calibration

Calibrate 30 mV input range (internal
reference source used for scaling the
input)

100mv

100 mV
calibration

Calibrate 150 mV input range (internal
reference source used for scaling the
input)

200mv

200 mV
calibration

Calibrate 300 mV input range (internal
reference source used for scaling the
input)

stroff

Strain offset

Calibrate the offset of the input circuit

v Lo

mV low

Calibrate low mV input (e.g. 0 mV)