Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual
21
Range (Switch 2)
The range switch adjusts the gain of the internal circuitry. Narrower 
ranges require more amplification. 
Switch
2
ON Range
Higher internal 
gain 
Turn this switch ON for the 
following input ranges: 15, 
30, ±15, or ±25 mV. 
OFF Range
Lower internal 
gain 
Turn this switch OFF for the 
following input ranges: 150, 
300, ±150, or ±250 mV. 
 Polarity (Switch 3) 
The polarity switch selects whether the range is unipolar and starts at 
zero (i.e. 0 to 30 mV) or bipolar and starts at a negative value, or below 
zero (i.e. -15 to +15 mV).  
Switch
3
ON Polarity
Range starts at 
zero (unipolar) 
Turn this switch ON for the 
following input ranges: 15, 
30, 150, or 300. 
OFF Polarity
Range starts 
below zero 
(bipolar) 
Turn this switch OFF for 
the following input ranges: 
±15, ±25, ±150, or ±250 mV. 
 Source (Switch 4) 
The source switch tells the ProVu whether the input is a strain gauge 
bridge (i.e. Figure 11), or it is a signal from a 2 or 4 wire transducer, or 
mV source (i.e. Figure 12) 
Switch
4
ON Source
Source is mV
input transducer
Turn this switch ON if the 
connected source is a 
transducer* 
OFF Source
Source is strain
gauge bridge
Turn this switch OFF if the 
source is a strain gauge 
bridge 
 
 
 
 
 
*Ratio should be set to NO in Setup Input Menu