Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual

21

Range (Switch 2)

The range switch adjusts the gain of the internal circuitry. Narrower
ranges require more amplification.

Switch

2

ON Range

Higher internal
gain

Turn this switch ON for the
following input ranges: 15,
30, ±15, or ±25 mV.

OFF Range

Lower internal
gain

Turn this switch OFF for the
following input ranges: 150,
300, ±150, or ±250 mV.

Polarity (Switch 3)
The polarity switch selects whether the range is unipolar and starts at
zero (i.e. 0 to 30 mV) or bipolar and starts at a negative value, or below
zero (i.e. -15 to +15 mV).

Switch

3

ON Polarity

Range starts at
zero (unipolar)

Turn this switch ON for the
following input ranges: 15,
30, 150, or 300.

OFF Polarity

Range starts
below zero
(bipolar)

Turn this switch OFF for
the following input ranges:
±15, ±25, ±150, or ±250 mV.

Source (Switch 4)
The source switch tells the ProVu whether the input is a strain gauge
bridge (i.e. Figure 11), or it is a signal from a 2 or 4 wire transducer, or
mV source (i.e. Figure 12)

Switch

4

ON Source

Source is mV

input transducer

Turn this switch ON if the
connected source is a
transducer*

OFF Source

Source is strain

gauge bridge

Turn this switch OFF if the
source is a strain gauge
bridge





*Ratio should be set to NO in Setup Input Menu