Shunt Calibration
The PD6100 is equipped with a means of simulating strain in a strain
gauge bridge circuit, via an included shunt resistor in the meter. This
technique is performed by enabling the “shunt resistor” switch (switch
5), which in turn shunts one leg of a connected strain bridge with a
EX+
COM
2
1
3
4
F4
SG+
SG-
INPUT SIGNAL
5
6
EX+
EX-
SG-
SG+
EX-
5-10 VDC
Power
Supply
+
-
EX+
COM
2
1
3
4
F4
SG+
SG-
INPUT SIGNAL
5
6
EX-
CONFIG
ON
4
5
3
2
1
3-Wire 100 mV
Transducer
EX+
SG-
SG+
Notes
Two-wire mV signals: Connect positive
to SG+ and negative to SG-
Four-wire transducer: Connect EX- to
negative power transducer terminal.