Model PD6100 Strain Gauge/Load Cell Meter Instruction Manual

23

NOTE:

Refer to Switch Configuration starting on page 20 for proper

switch positioning for the following wiring configurations.

Figure 11. Strain Gauge Powered by External Supply

Figure 12. mV Transducer Input Connections

Shunt Calibration

The PD6100 is equipped with a means of simulating strain in a strain
gauge bridge circuit, via an included shunt resistor in the meter. This
technique is performed by enabling the “shunt resistor” switch (switch
5), which in turn shunts one leg of a connected strain bridge with a

EX+

COM

2

1

3

4

F4

SG+

SG-

INPUT SIGNAL

5

6

EX+

EX-

SG-

SG+

EX-

5-10 VDC

Power

Supply

+

-

EX+

COM

2

1

3

4

F4

SG+

SG-

INPUT SIGNAL

5

6

EX-

CONFIG

ON

4

5

3

2

1

3-Wire 100 mV

Transducer

EX+

SG-

SG+

Notes

Two-wire mV signals: Connect positive
to SG+ and negative to SG-

Four-wire transducer: Connect EX- to
negative power transducer terminal.