beautypg.com

Orientation/angle angles, cones, Orientation/co-planarity planes, Runout/circular runout circles, arcs – HEIDENHAIN ND 1400 User Manual

Page 95

background image

81

7

Tolerancing

Orientation/Angle angles, cones

Angle tolerances can be applied to angles and cones. The tolerance screens are identical, and are used to

specify the acceptable deviation from a nominal angle.

Enter the allowed angle limits into

the Limit data fields as +/- ranges

or as absolute limits. Touch the

Limit data fields to toggle between

+/- ranges or absolute limits.

Orientation/Co-planarity planes

Co-planarity tolerances can be applied to planes and are used to specify

the acceptable deviation from a plane perfectly parallel to a reference

plane.

Touch the Parallelism icon for co-planarity tolerances. Enter the ac-

ceptable co-planarity deviation into the Tol Zone data field, and then

enter the number of the reference feature for the parallelism tolerance

into the Ref Feature data field.

Runout/Circular runout circles, arcs

Circular runout tolerances can be applied to circles and arcs. The tol-

erance screens are identical, and are used to specify the acceptable

deviation from a circle or arc around the center of the reference feature

axis.

Enter the acceptable runout deviation into the Tol Zone data field and

then enter the number of the reference feature for the tolerance into the

Ref Feature data field.

Tolerance Types