122

QC-300 Series User’s Guide

5 Enter the MZ Offset value noted earlier in step 2.

6 Zero the DRO for the axis being calibrated at the part zero position of the

standard and enter zeros into the Standard and Observed data fields for Station

zero.

7 Measure the standard, increment-

ing the Station number and entering

the Standard (certified) and Observed

(measured) values into the correspond-

ing data fields for each segment.

This example shows measurements

across 150 mm on the X-axis, and the

SLEC screens for entering the six seg-

ments of standard and observed data to

provide error correction.

Reminder

Segments are straight lines on the

deviation graph. Only the beginning

and end points of each straight line

segments must be entered into SLEC

data fields.

8 Touch the Enabled choice field to toggle it On.

Station 1 of 6

Station 2 of 6

Station 3 of 6

Station 4 of 6

Station 5 of 6

Station 6 of 6