Setup

2 Make sure the Enabled choice field is toggled to Off, and erase any existing

data in all the data fields using the methods described earlier in this chapter. Sta-

tion 0 should be displayed with zeros in the Standard, Observed and MZ Offset

fields when this step is complete.

3 The length of the calibration standard should cover as much of the axis range

of motion as possible. Measure the standard and enter the Standard (certified) and

Observed (measured) values into the corresponding data fields.

4 Touch the Enabled choice field to toggle

error correction On.

SLEC Screen