40

QC-300 Series User’s Guide

Part skew alignment

When the part is misaligned (twisted) in the X-Y plane, errors can be generated

during the measurement process. These errors can be eliminated by performing a

skew alignment on the part.

The skew alignment includes a precise measurement of the part misalignment.

Once the misalignment is known to the system, subsequent feature measurements

are automatically compensated to eliminate misalignment errors. Measurement

data in the DRO will reflect measurements of a perfectly aligned part.

To skew the part:

1 Touch the Measure tab and then touch the measure line icon.

2 Measure a line along the reference edge of the part. The X-axis

edge is used in this example.

3 Touch the Align button to

perform a skew on the reference

edge.

The skew line feature will now be shown in blue and the reference angle will be zero degrees.