Ports screen – HEIDENHAIN ND 1400 User Manual
Page 152

138
QC-300 Series User’s Guide
Ports screen
The Ports screen contains fields for configuring the
RS-232 serial communication port. The RS-232
communication port is used to send data to comput-
ers. RS-232 port settings can be changed to match
those of the receiving system.
Baud rate
Touch the Baud Rate choice field repeatedly to cycle
to the desired serial data rate. Baud rates from 1,200
to 115,200 are
available.
Word length
Touch the Word Length choice field to toggle between data word
lengths of 7 bits and 8 bits.
Stop bits
Touch the Stop Bits choice field to toggle between 1 and 2 stop bits at the
end of the data word.
Parity
Touch the Parity choice field repeatedly to cycle to the desired parity error
checking. Choices include None, Odd and Even.
EOC delay
A delay can be inserted at the end of each character to satisfy the communication requirements of receiv-
ing systems. Touch the EOC data field and enter the desired
delay in milliseconds using the number keys.
EOL delay
A delay can be inserted at the end of each line of characters to satisfy the communication requirements
of receiving systems. Touch the EOL data field and enter the
desired delay in milliseconds using the number keys.