beautypg.com

HEIDENHAIN ND 1400 User Manual

Page 142

background image

128

QC-300 Series User’s Guide

Probe hit starts measure magic

Measure magic automatically selects a feature type, such as line or circle, based on the geometric orienta-

tion of probed data points. Measure magic can be started

manually by touching the measure magic icon prior to a mea-

surement, or started automatically when the first new probe

contact of a measurement occurs, thereby saving the user the

time required to press a panel button. Enable this automatic

function by toggling the Probe Hit Starts Magic choice field

to Yes.

Auto save UCS

A new user coordinate system (UCS) is the measurement reference frame established each time the user

creates a new datum. The new UCS is temporary until it is saved or replaced by another new UCS. The

UCS can be saved manually by touching the UCS icon in the top right corner

of the screen, or can be saved automatically each time a new datum operation

is performed by the operator. Enable this automatic function by toggling the

Auto save UCS choice field to Yes.

Distances

Distance measurements result in X, Y and L (vector) displacements. The Y-axis displacements can be

shown as signed or absolute (positive only) values. When signed values are specified, axis displacements

from right to left and from top to bottom are negative. Specify signed distances by toggling the Distances

choice field to Signed.

Enabling and configuring point filtration

Point filtration extracts aberrant data points from the total population of data during the least squares best fit

calculations of arc, circle and line features. Points are extracted when they exceed the specified error limit

(quantization factor) and fall outside the specified standard deviation range (Sigma factor). The filtration

process ends when all remaining data points satisfy the quantization or standard deviation requirement, or

when the minimum percentage of retained data points (proportion factor) is reached. Extracted data points

are highlighted in yellow in the View window. The last points retained are highlighted in red.