Reports are not printed or are incomplete – HEIDENHAIN ND 1400 User Manual

Page 161

background image

147

11

Problem Solving

Values displayed on the LCD screen are incorrect (continued)

Possible cause Possible solutions

Wrong in/mm unit of measure

1 Specify the correct units of measure in the

Encoders setup screen

Wrong encoder reference mark is specified

1 Correct the reference mark type in the

Encoders setup screen

Error correction is required to compensate 1 Perform LEC or SLEC error correction in the

encoder inaccuracies SLEC setup screen

Touch probe is not calibrated 1 Perform a probe teach operation as described

in Chapter 5: Probes

Touch probe configuration settings 1 Change the touch probe settings in the Probe

are incorrect setup screen

Reports are not printed or are incomplete

Possible cause Possible solutions

Printer is not supported 1 Browse to www.metonics.com for a list

of supported printers

USB cable is not firmly connected 1 Connect or replace the printer cable

or is damaged

Printed report setup is incorrect 1 Correct the report setup in the Print

setup screen

Symptoms, Probable Causes & Solutions