beautypg.com

Seaward Supernova User Manual

Page 85

background image

Leakage

75

Leakage

Range...........................................................................0.1mA - 10.0mA

Resolution.............................................................................0.01mA

Accuracy................................................ +/- 5% of reading, +/- 2 digits

Pass Levels.....................0.25mA, 0.5mA, 0.75mA, 1.5mA, 2.25mA,

2.5mA, 3.0mA, 3.5mA, 9.9mA

Touch Leakage

Range.............................................................................0.1mA - 2.0mA

Resolution.............................................................................0.01mA

Accuracy.............................................. +/- 10% of reading, +/- 2 digits

Pass Levels...................................................0.25mA, 0.5mA, 1.0mA

Power Tests

Socket......Selectable 115V BS4343, BS546 / 230V BS 1363 socket.

Measured Voltage................................................................. 90 - 300V

Indication only.

Measured Load ....................................................................... 0 - 4kVA

Indication only.

IEC Lead Test

Test Voltage *…40Vac or 250Vdc depending on hardware revision

Detects..................................................... Open, Short, Good, Reverse

* - Test stimulus based on mains supply of 230V, varies with supply
voltage.