Seaward Supernova User Manual
Page 29

How to perform Automatic Tests
19
Sequence, a list of options for that field will be displayed:-
Use the cursor left and right keys to highlight the description, then
use the cursor up and down keys to highlight the required setting.
Use the cursor left and right keys to change the highlighted field.
Use the Green button key to save the settings and return to the
Default Sequence Editor. Repeat this process for each field as
appropriate.
Press the red button to abort changes made to the settings and
return to the Default Sequence Editor.
To start the test sequence select PERFORM TEST.
Use Testcodes
This option bypasses the need to edit a specific test sequence,
allowing direct entry of a Testcode through the keyboard or a barcode
scanner.
Use the cursor keys to highlight the Testcode to be changed. Press
Green button to select that digit. Use the keyboard to change the
value of the digit. Press Green button to confirm the change. Press
the Red button to abort changes to the selected Testcode.
Use a barcode scanner to enter a Testcode
---------------ENTER TESTCODES-------------------------
OUTPUT LIMIT DUR NUM
SOCKET
230V
VISUAL
YES
1
EARTH BOND
4A 0.10
Ω 5 1
INSULATION
500V 0.50M
Ω 5 1
SUB LEAKAGE
SKIP
FLASH
SKIP
LEAKAGE
SKIP
TOUCH LEAKAGE
SKIP
POLARITY
SKIP
TESTCODE
1:
I2F8008032
TESTCODE
2:
R222111111
PERFORM TEST