Sequence Warning Faults

61

The Supernova is faulty and should be returned to Seaward for
Repair

Sequence Warning Faults

If you enter a testcode which involves tests the Supernova cannot
perform then you may see one of the following warnings.

--------------------WARNING--------------------
THE SELECTED TESTCODE SPECIFY
THAT A 200mA (DC) EARTH BOND
TEST SHOULD BE PERFORMED
THIS IS NOT POSSIBLE ON THE
SUPERNOVA
A 100mA (AC) TEST WILL BE
PERFORMED. PRESS START KEY TO
CONTINUE

Safety Tests during operation

The Tester performs self-tests during normal operation. An internal
earth bond test and a low voltage test are performed prior to applying
mains power to any EUT (including Load/Leakage and Touch
Leakage tests). If the internal earth bond test is failed, the following
message will be displayed:-

INTERNAL RELAY FAULT

CANNOT PROCEED !!

If this message appears, the Supernova is faulty and requires repair.

If the low voltage test fails, then either of the following warnings will
be displayed:-

--------------------WARNING--------------------

APPLIANCE MAY BE SHORT CIRCUIT
PRESS THE ENTER BUTTON IF SAFE
PRESS THE RED BUTTON TO ABORT