Sequence warning faults, Safety tests during operation – Seaward Supernova User Manual
Page 71

Sequence Warning Faults
61
The Supernova is faulty and should be returned to Seaward for
Repair
Sequence Warning Faults
If you enter a testcode which involves tests the Supernova cannot
perform then you may see one of the following warnings.
--------------------WARNING--------------------
THE SELECTED TESTCODE SPECIFY
THAT A 200mA (DC) EARTH BOND
TEST SHOULD BE PERFORMED
THIS IS NOT POSSIBLE ON THE
SUPERNOVA
A 100mA (AC) TEST WILL BE
PERFORMED. PRESS START KEY TO
CONTINUE
Safety Tests during operation
The Tester performs self-tests during normal operation. An internal
earth bond test and a low voltage test are performed prior to applying
mains power to any EUT (including Load/Leakage and Touch
Leakage tests). If the internal earth bond test is failed, the following
message will be displayed:-
INTERNAL RELAY FAULT
CANNOT PROCEED !!
If this message appears, the Supernova is faulty and requires repair.
If the low voltage test fails, then either of the following warnings will
be displayed:-
--------------------WARNING--------------------
APPLIANCE MAY BE SHORT CIRCUIT
PRESS THE ENTER BUTTON IF SAFE
PRESS THE RED BUTTON TO ABORT