Seaward Supernova User Manual
Page 62

52
Sub Leakage
This option allows you to set the Sub Leakage value indicated 10%,
6% or 0% high, depending upon national legislation or best practice.
Power Polarity
The Power Polarity option will allow the tester to perform the power
tests in one of three ways.
Input Polarity:- perform the power tests with the same mains polarity
as the Supernova is powered.
Auto-corrected:- perform the power tests with the mains correct
polarity even if the Supernova is powered from a socket which has
the mains polarity reversed.
Input & reversed:- this option will perform two power tests for every
one selected, one with the correct mains polarity and one with the
reverse mains polarity.
On Test Failure
This gives you the option of having a menu for a failure within an
automatic sequence. This menu will give you the option of retrying a
failed test or appliance.
Roll Printer
This allows the user to define the thermal printer to be used with the
Supernova. Seawards recommended option is “SEAWARD
(283A954)”
Date & Time
This allows the user to password protect the date and time. An
existing password must be setup before this option can be used.
-----------------SYSTEM CONFIGURATION----------------
SUB LEAKAGE
0%
POWER POLARITY
INPUT & REVERSED
ON TEST FAILURE
END APPLIANCE
ROLL PRINTER
A-P860B
DATE AND TIME
UNPROTECTED
USER LIST
UNPROTECTED
APLNCE & CMNTS
FULL LENGTH
OK