Earth Bond Test

73

Chapter 7 Specifications

Earth Bond Test

Test Voltage * .....................................................6V nominal (no load)

Test Current *.....................4A, 10A, 25A selectable (into s/c load)

Range .......................................................................... 40m

Ω - 19.99Ω

Resolution...............................................................................0.01

Accuracy................................................ +/- 5% of reading, +/- 2 digits

Pass
Levels
...........................................................................selectable

Misc.............................................4-wire measurement, floating earth

Earth Screen Test

Test Voltage * ............................................. 100mV nominal (no load)

Test Current *...............................................100mA (into s/c load)

Range............................................................................ 40m

Ω - 19.99Ω

Resolution...............................................................................0.01

Accuracy..............................+/- 5% of reading, +/- 2 digits upto 5.00

Pass
Levels
...........................................................................selectable

Misc.............................................4-wire measurement, floating earth