Shield termination, Customer interface cable requirements – MicroE Mercury II 6000V User Manual
Page 19
Page 19
Section C - Reference
Customer Interface
Cable Requirements
Fold braided shield back over jacket. Example shows double-shielded cable.
Dimensions shown are for illustration only.
7.6 [.30]
28.7 [1.13]
Aluminum Polyester Shield not to
be exposed in this area.
Do not twist.
Jacket
Aluminum
Polyester
Shied
Braided
Shield
C4.2
Shield Termination:
The customer's cable shield should be in 360° contact
with the connector shroud and the connector shell to
provide complete shielding. The connector shell should
be metal with conductive surfaces. Suggested metal
connector shells for use with Mercury
II
™
encoders:
AMP 748676-1 or equivalent; where the dash
number is dependent on the customer's outside cable
diameter. The shield should be terminated as
illustrated in the following diagram.