Grounding instructions – MicroE Mercury II 6000V User Manual
Page 15
Page 15
Section C - Reference
Grounding Instructions -
Mercury
II
™
6000
Sensor mounted to a surface that is
grounded through bearings or a poorly-
grounded surface, or mounted to a non-
conducting surface
C
C1
1..2
2..1
1
15-pin D-sub connector grounding: The
encoder's connector shell must be in intimate,
electrically conductive contact with the
customer-supplied mating connector, which
must be connected to the controller's ground.
If a customer-supplied shielded cable
connects the encoder to the controller, then
the outer shield on the customer-supplied
cable must be connected to the controller's
ground. The controller must be grounded to
earth at the point of installation.
C
C1
1..2
2..2
2
The encoder sensor must be mounted so that
it is electrically isolated from ground.
5 Volts
0 Volts
POWER
SUPPLY