Interface cable requirements, Customer interface cable requirements – MicroE Mercury II 6000V User Manual
Page 18

Section C - Reference
Page 18
Customer Interface
Cable Requirements
Mercury
II
™
6000
Signal
Twisted Pair
A+
Pair 1
A-
B+
Pair 2
B-
Index+
Pair 3
Index-
Left Limit +
Pair 4
Left Limit -
Right Limit +
Pair 5
Right Limit -
+5V
Pair 6
GND
C4.0
Customer cables that interface to Mercury
II
™
series
encoders must have the following characteristics:
• Twisted pair signal wiring.
• Characteristic impedance of 100-120 ohms.
• Sufficient wire gauge to meet the minimum
voltage requirement at the encoder, for example
24AWG gauge wire for a 2m length cable.
Examples of acceptable cables with 24AWG
gauge wire and 4 twisted pairs are Belden
9831, 8104, and 9844 or other
manufacturer’s equivalents.
• Single shield cable with a minimum of 90%
coverage. Note that a double shielded cable
may be required in high-noise applications.
C4.1
Signal Wiring
Each differential signal should be connected to
a corresponding twisted pair as follows: