Guralp Systems CMG-DM24 User Manual
Page 41
![background image](https://www.manualsdir.com/files/788444/content/doc041.png)
Operator’s Guide DM-24 Digitiser
Issue G January 2003
39
SENSOR MASS CONTROL
Command
Syntax
Comments
LOCK
LOCK
UNLOCK
UNLOCK
CENTRE
CENTRE
The 3 commands LOCK,
UNLOCK, and CENTRE are
simply entered at the command line.
While they are executing the system
displays the mass positions for the 3
components updated once per second
(3 seconds to cycle through the 3
components). The mass positions are
measured to 16 bits so when the
masses are locked the displayed
readings will be in the region of 32k
or -32k. When the masses are
correctly centred the readings should
be less than 1000 counts.
AUXILIARY (MUX) CHANNELS
Command
Syntax
Comments
SET-CONFIG
SET-CONFIG
See digitiser output
programming earlier
If you are programming the digitiser using the SCREAM terminal window or a
standard terminal program, you must issue the RE-BOOT command to restart the DM.
Command
Syntax
Comments
RE-BOOT
RE-BOOT
no parameters
The command prompts the
operator to confirm this
operation with ‘y’. The
system will then
automatically reset after a
delay of about 2 s.